Dual Channel Stim Unit; suitable for urinary or rectal incontinence to educate and motivate patients ; for daily use to help retrain the pelvic floor. Use with Periform, Veriprobe, Novaty, Anuform or Surface electrodes sold separately.
Key Benefits - Download Brochure (PDF)
Dual channel PelviTone device has all the features of Continence device, plus:
- Advanced multi-phase programmes where treatment is split into time phases with different stimulation parameters;
- Stronger symmetrical stimulation with the reduced Pulse Width : 50 – 330 μS.
- Three Custom programmes plus 11 Preset Continence programmes.
- Lock mode function to measure the home compliance (mA and time used)
- NeuroTrac ™ VeriProbe - The Vaginal probe for EMG and Electrostimulation.
- Suitable for urinary or rectal incontinence
- Digital dual channel
- To educate and motivate patients
- Set work/rest times and repetitions
- 11 built in programmes
- 3 customisable 5 segment programmes
- Records statistics:- Work/rest times /Usage Onset/muscle release times
- User friendly
- Can be used in pocket or upright on flat surface
PLEASE NOTE THAT THE NEUROTRAC™ PRODUCTS ARE NOT FOR SALE OR USE IN THE USA or CANADA
Specification
- Dual channel: individually isolated circuits.
- Amplitude: 0-90 mA into 500 Ohm load; actual mA will tend to be less than indicated due to electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 86 mA, at 1500 Ohms load the maximum will be limited to 65 mA.
- Type: Constant current,
- maximum output voltage 180 Volts +10 / -30 Volts
- Waveform: Symmetrical, rectangular bi-phasic with zero DC
- current.
- Selectable pulse width: 50 μS – 330 μS [2% accuracy].
- Pulse Rate selection: in the continuous mode 2 – 100 Hz
- [2% accuracy].
- Time duration of the treatment selectable: 1 minute to 90 minutes.
- Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2
- volts the battery symbol will flash on/off once every second.
- Open Electrode Detect: If an open circuit is detected at the
- output of channel A or B the output current will be reset at zero.
- Ramp up time 0.1 - 9.9 seconds.
- If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn
- on.
- Physical dimensions: 134 x 69 x 29.7 mm.
- Weight: 0.18 KG with battery.
- Environmental Conditions for use:
- +10 to +30 degrees Centigrade. 0-90% Humidity.
- Environmental conditions for storage & transport:
- -10 to +50 degrees Centigrade. 0-90% Humidity.