Phase 3

Defibrillator / Pacer Analyser - Real-time waveform capture and data transfer to your PC

Phase 3 is the first Defibrillator/Pacer Analyser specifically designed to test Pulsed Multiphasic defibrillators. Test Monophasic, Biphasic and Pulsed Multiphasic Defibrillators, AEDs and Transcutaneous Pacers with better than 1% accuracy on either line or battery power.

Battery provides up to 24 hours or continuous operation and detachable paddle plate provides easy hands-free testing.

Test All Defibrillators

Including monophasic, biphasic, and pulse biphasic waveforms

Test AEDs and Pacers

32 Auto Sequences

Save you time, money and manpower

Lightweight + Battery Operated


Testing capabilities
USB Communication capability allows real-time waveform capture and efficient data transfer to your PC. Capture high-resolution signal data to a PC running Phase3pc software. Store up to 80 test records and up to 10 defibrillator discharge waveforms as Wave Records.

Use Phase 3pc companion software to create and edit auto sequences, download and store test data and wave graphics to your PC. Phase3pc companion software is included as a standard accessory.

Versatility
Phase 3 tests Monophasic, Biphasic and Pulsed Multiphasic Defibrillators, AEDs and Transcutaneous Pacers with better than 1% accuracy on either line
or battery power.

Portability
Detachable Paddle Plate provides easy hands-free testing. Battery provides up to 24 hours of continuous operation.

USB Connectivity
USB Communication capability allows real-time waveform capture and efficient data transfer to your PC.

Real-Time Wave Capture
Capture high-resolution signal data to a PC running Phase3pc software immediately after discharge.

Test Record Storage
Phase 3 provides storage for up to 50 Test Records and up to 10 defibrillator discharge waveforms as Wave Records.

Variable Load Module
Our optional, exclusive VLM allows testing defibrillators at maximum energy with selectable loads from 25 to 175 ohms, in accordance with AAMI DF-80 and IEC 60601-2-4. High-resolution wave graphics and test data are obtained with Phase 3VL software — a standard accessory to the VLM.

Phase3pc Companion Software
Included as a standard accessory, Phase 3pc allows you to create and edit auto sequences, download and store test data and wave graphics to your PC.

Defibrillator Energy, Charge Time and Cardioversion Automated
External Defibrillator (AED) Performance
ECG Monitor Performance
Pacer
Sensitivity, Refractory Periods and Noise Immunity
Pacer Pulse
Characteristics
Energy Measurement, General
Load resistance: 50 ohms ±1%, non-inductive
ECG amplitude at defib pads: 1 mV QRS
WAVEFORM (oscilloscope) Output
High Range: 1000:1 amplitude attenuation
Low Range: 200:1 amplitude attenuation
Waveform Playback: 200: 1 time base expansion
First Phase Pulse Width: 0.1 – 58 msec
Second Phase Pulse Width: 0.1 – 58 msec
Interphase Delay: 0.1 – 58 msec
Tilt: 0 – 99.9%
Modulation Frequency: 1000 – 8000 Hz
Modulation Duty Cycle: 0 – 99.9%
Test Pulse: 46 Joules ±10%
Defibrillator High Range Energy Test
Energy Measurement: 0.0 to 600.0 Joules (±1% ±2 LSD)
Voltage Measurement: 0 to 5000 Volts (±1% ±2 LSD)
Current Measurement: 0.0 to 100.0 Amps (±1% ±2 LSD)
Pulse Width Measurement: 0.5 to 58.36 msec (±1% ±2 LSD)
Trigger Level: 80 Volts
Playback Amplitude: 1 mV per 1000 volts on Lead II;
1 mV per 2000 volts at defib pads
Test Pulse: 126 Joules ±10%
Defibrillator Low Range Energy Test
Energy Measurement: 0.0 to 50.0 Joules (±1% ±2 LSD)
Voltage Measurement: 0 to 1000 Volts (±1% ±2 LSD)
Current Measurement: 0.0 to 20.0 Amps (±1% ±2 LSD)
Pulse Width Measurement: 0.5 to 58.36 msec. (±1% ±2 LSD)
Trigger Level: 16 Volts
Playback Amplitude: 1 mV per 200 volts on Lead II;
1 mV per 400 volts at defib paddles
Defibrillator Charge Time Test
Charge Time Measurement: 0.0 to 99.9 seconds (±1 LSD)
Defibrillator Cardioversion Test
Sync Delay Measurement: -200 to +800 msec. (±1 LSD)
Delay Target: 20 to 65 msec window when enabled
Sync Point: Selectable, peak of ECG Q or R wave
AED Performance Test
Test Method: Verify AED shock advisory for specified arrhythmia
Pacemaker Pulse Test
Pulse Amplitude Measurement: 4 to 250 milliamps, all loads (±1% ±1 LSD)
Pulse Rate Measurement: 20 to 220 PPM (±1% ±1 LSD)
Pulse Width Measurement: 0.5 to 58.36 msec. (±1% ±2 LSD)
Test Load Range: 50 to 1600 ohms, in 50 ohm steps
WAVEFORM output: 50 milliamps per volt, all loads
Measurement Methods: Average, leading edge, trailing edge, peak
Test Pulse: 145 mA ±10%
Pacemaker Noise Immunity Test
Test Waveform: 50Hz or 60Hz sine wave
Noise Amplitude Range: 0.18 to 10.00 mV peak-to-peak
Noise Amplitude Resolution: 0.139 mV
Pacemaker Sensitivity Test
Test Waveform: Square (SQR), Triangle (TRI) or Haversine (SSQ) pulse
Waveform Width: 10, 25, 40, 100, or 200 msec.
Amplitude Range: 0.00 to 3.00 mV peak
Pacemaker Refractory Period Test
Paced Refractory Period (PRP): 50 to 750 msec. (±1 LSD)
Sensed Refractory Period (SRP): 50 to 750 msec. (±1 LSD)

ECG Simulator

Performance Test Waveforms
DC Pulse, 4 seconds, Square Wave, 2 Hz, Triangle Wave, 2 Hz Sine
Wave @ 0.1, 0.5, 10, 20, 40, 50, 60, 70, or 100 Hz
Normal Sinus Rhythm
30, 60, 90, 120, 150, 180, 240 or 300 BPM
Cardioversion, Shock Advisory and AED Test Waveforms
Atrial Fibrillation, Coarse
Atrial Fibrillation, Fine
Asystole
1 (random, low-frequency baseline fluctuation)
Asystole 2 (flat line/zero
volts) Supraventricular Tachycardia (SVT-140)
VTACH @ 140, 160, 190 BPM
Torsades de Pointe @ 200 BPM
Coarse Ventricular Fibrillation (CVF) / Fine
Ventricular Fibrillation (FVF)
Square Pulse 1ms to 60 BPM
Arrhythmia Simulations
Second Degree A-V Block, Premature Atrial Contraction (PAC), R-on-T
PVC
Right Bundle Branch Block (RBBB), Premature Ventricular Contraction (PVC)
Multifocal
PVC, Run of 5 PVC, Bigeminy, Trigeminy
Pacemaker Test Waveforms
SQR (square) Pacer Trigger, width = 2, 25, 40, 100 or 200 msec
TRI
(triangle) Pacer Trigger, width = 10, 25, 40, 100 or 200 msec
SSQ (haversine)
Pacer Trigger, width = 10, 25, 40, 100 or 200 msec
Performance Specifications
Output Level: Selectable, 1 mV, 2 mV or 0.5 mV into ECG Lead II
Impedance: 500 ohms (±0.2%), Amplitude: ±2%

Non-Volatile Memory

Data Capacity
50 Test Records, 10 Defibrillator Waveform Records, 32
Autosequences
Test Record Content
Device ID, Time/date of test, Test type (Manual or Auto), Device
type (defib. or AED)
Up to 10 defibrillator energy tests (or 32 AED energy
tests)
1 defibrillator charge time test
Up to 4 defibrillator
cardioversion tests
Up to 12 ECG performance tests
Up to 10 pacer pulse
tests
1 pacer noise immunity test
Up to 2 pacer sensitivity tests
Up to
2 pacer refractory period tests
Interface
LCD (5.2″ x 1.5″; 40 characters x 8 lines text; 240 x 64 pixel
graphics) Defibrillator Input: Molex 42820-3212
Pacemaker Input: 2 x
safety-style banana jack (red (+)/black (-))
ECG Simulator Outputs: 10 x
safety banana jack (RA; RL; LA; LL; V1-V6)
Defibrillator/Pacer Waveform
Output: 1/8″ mono phono jack
High-Level ECG Output: 1/8″ mono phono jack
USB
Port: Type “B”, USB 1.1 or USB 2.0 compatible.
Serial (RS-232) Port: DB9
Male, RS-232C, bi-dir, CTS handshaking, 9600 baud, 8-N-1 Keyboard Port: PS/2
(6-pin miniDIN female)
Power Supply
Internal 12.5V/1.4A-h NiCad, 24 hours of use between charges
Environment
15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II
Dimensions
9.5″ W x 8″ H x 5.5″ D (24cm W x 20cm H x 14cm D)
Weight
3 lbs. (1.4 kg)

All specifications subject to change without notice.

Support Information

Helpdesk:
Head Office:
  • Address
    • de Smit Medical Systems Ltd, Unit 15 Falcons Gate, Dean Road, Bristol, BS37 5NH, UK
  • Operating Hours
    • Monday to Friday - 9am-5pm (Excluding bank holidays)
Service Centre:
  • See Head Office above
  • Please contact the Service Centre regarding any equipment servicing/repair related queries or telephone troubleshooting
Trained Personnel:
  • 7 x Sales technicians
  • 3 x Medical engineers at Head Office
    • All trained to manufacturer standard
Turnaround:
  • Warranty Repairs
    • 72 working hours response/resolver call-out for warranty repairs and service plans
  • Non-warranty Repairs
    • 72 hours subject to authorisation by Customer for paid repair
Spare Parts:
  • Spare parts for MCube/Datrend products are available direct from de Smit Medical Systems Ltd
  • Next working day delivery if order received by 2pm on any given working day (subject to availability)

Five-Year Warranty on Datrend legacy Products

All Datrend legacy products are designed and manufactured under an ISO 13485 registered Quality System and a carry one (1) year from the date of original purchase.

This warranty will be automatically extended on a yearly basis, to a maximum of five (5) years from the date of original purchase, provided that calibration is performed on an annual basis by de Smit Medical Systems.

Please contact de Smit Medical Systems Ltd for further information about this product including;

  • Full product information / specification
  • Arranging a demonstration / evaluation
  • Pricing details
  • How to place an order
Call +44 (0)845 345 4226, or email us via our contact form.