vPad-RF

The smallest full featured Electrosurgical Unit analyser

Introducing vPad-RF, the smallest full featured analyser for testing Electrosurgical Units (ESUs). Built on Datrend’s revolutionary Vision Pad platform, each unit comes with a built-in HF Digital Oscilloscope and Spectrum Analyser.

vPad-RF is the smallest full featured analyser for testing Electrosurgical Units (ESUs). Built on our Vision Pad platform, vPad-RF has all the functionality you have come to expect from products built on this platform. What really makes this product unique is its built-in HF Digital Scope and Spectrum Analyser.

Full Featured ESU Analyser

User Programmable

Save Time, Money and Manpower!

Automate Test Sequences

Easy-to-read graphics and reports

Eliminate Human Error


  • Digital signal processor (DSP) enables accurate testing of all ESU makes and models, continuous and pulsed-waveforms
  • Integration of ESU testing with HF digital storage oscilloscope and spectrum analyser functions provides three pieces of test equipment in one
  • High powered RF load provides settings required for all IEC 60601-2-2 tests, including Covidien ForceTriad maintenance procedures
  • Wide measurement range – up to 8.5A
  • Pre-Configured automated PM routines for most major manufacturers
  • Works with vPad safety analysers and automation apps.
Please contact de Smit Medical Systems Ltd for further information about this product including;

  • Full product information / specification
  • Arranging a demonstration / evaluation
  • Pricing details
  • How to place an order
Call +44 (0)845 345 4226, or email us via our contact form.
HF Power (RMS)
Range 0 – 999.9 W (as derived from HF current)
Accuracy ±(3% of reading + ½W)
HF Current (RMS)
Range 0 – 5A, int. load bank 0 – 8.5A, ext. load
Accuracy ±(1% of reading + 5mA)
Voltage (RMS)
Range 0-999.9 V (as derived from HF current)
Accuracy ±(1% of reading + 1V)
Voltage (Peak)
Range 0 – 9999 V
Accuracy ±(1% of reading + 10V)
Crest Factor
1.0 – 999.9 (as derived from peak/RMS ratio)
Instrument RF Bandwidth (-3dB)
250 Hz – 12 MHz
Isolation, PE to MD
10 kV
Variable Load
Range 0 – 5115 ohms
Resolution 5 ohms
Accuracy ±1% + 0.5, -0 ohms
Power, 50% duty cycle 200W @ 5 ohms
400W @ 10 ohms
600W @ 20 ohms
600W @ 40 ohms
600W @ 80 ohms
600W @ 160 ohms
600W @ 320 ohms
600W @ 640 ohms
530W @ 1280 ohms
260W @ 2560 ohms
Operational Duty Cycle
Up to 100% in manually activated tests
50% in automated
footswitch activation, from 1 to 20 seconds
Pulsed ESU
Measurements Pulse On-time, Off-time & Duty Cycle
Timing Range 0.1 to 6500msec
Resolution 0.1msec
Accuracy ± 0.2msec
Pulse Duty Cycle 0.1 to 99.9%
RF Leakage Load
Resistance 2 x 200 ohms
Accuracy ±1%
Power 100% duty 16W
REM / CQM Test Resistance
Range 0 – 1023 ohms
Resolution 1 ohm
Accuracy ±1% + 1, -0 ohms
HF Digital Oscilloscope
Vertical Ranges 0.125, 0.25, 0.75, 1.5, or 3A / division
Vertical Range Selection Manual or Automatic
Horizontal Ranges 1.25, 2.5, 5, 10, 20 or 40 µsec / division
Trigger Level Adjustable, +/- edge
HF Spectrum Analyser
Vertical Ranges Normalized dB or linear
Horizontal Ranges 62.5, 125, 250, 500, 1000, 2000 kHz / division
Frequency Resolution 1.22, 2.44, 4.88, 9.77, 19.5, 39.1 kHz
User Interface
Android 10” Tablet providing Graphical User Interface
WiFi & Bluetooth communication 12 Gb memory (or more – optional)
Power
120 Volt power outlet 90 to 132 V ac rms,
47 to 63 Hz, 0.6 A max.
230 Volt power outlet 180 to 264 V ac rms,
47 to 63 Hz, 0.4 A max.
Size 8.5 x 11.0 x 13 in.
Weight 18 lb.
Environment 15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II

Support Information

Helpdesk:
Head Office:
  • Address
    • de Smit Medical Systems Ltd, Unit 15 Falcons Gate, Dean Road, Bristol, BS37 5NH, UK
  • Operating Hours
    • Monday to Friday - 9am-5pm (Excluding bank holidays)
Service Centre:
  • See Head Office above
  • Please contact the Service Centre regarding any equipment servicing/repair related queries or telephone troubleshooting
Trained Personnel:
  • 7 x Sales technicians
  • 3 x Medical engineers at Head Office
    • All trained to manufacturer standard
Turnaround:
  • Warranty Repairs
    • 72 working hours response/resolver call-out for warranty repairs and service plans
  • Non-warranty Repairs
    • 72 hours subject to authorisation by Customer for paid repair
Spare Parts:
  • Spare parts for Mcube/Datrend products are available direct from de Smit Medical Systems Ltd
  • Next working day delivery if order received by 2pm on any given working day (subject to availability)

Two-Year Warranty on Datrend vPad Products

All Datrend vPad Products are designed and manufactured under an ISO13485 registered Quality System and carry a one (1) year warranty from the date of original purchase. This warranty will be automatically extended to two (2) years from the date of original purchase, provided that calibration is performed on an annual basis by a de Smit Medical Systems.