vPad-IN

Next generation Infant Incubator & Radiant Warmer testing system

Introducing the vPad-IN, the next generation infant incubator and infant radiant warmer testing system that meets the requirements of the IEC Standards for incubators (IEC 60601-2-19) and radiant warmers (IEC 60601-2-21).

vPad-IN has all the functionality and scalability that you have come to expect from Vision Pad Technology products. Parameters are sensed inside the incubator environment and results are transmitted wirelessly to a 10″ Android tablet for analysis and display. Test parameters and test protocols can be modified without disturbing the equilibrium of the heated environment, thereby minimising test times.

Wirelessly Control Tests

Changes to test can be made without disturbing heat environment

IEC 60601-2-19 and 2-21

The ONLY Incubator analyser to test to IEC 60601-2-19 and 2-21. Follows Edition 1 and Edition 2 of Standard

vPad Platform

Built on the Vision Pad platform to give you greater flexibility and scalability

Automated Measurements

Measures all parameters automatically, following user defined protocols

Compatible with closed, forced air convection incubators and open, radiant infant warmers (optional)

  • Simultaneous measurement of humidity, airflow, sound, and 6 different temperatures
  • The only Tester available to test in compliance with IEC 60101-2-19 and IEC 60601-2-21
  • All sensors are independently movable, to be placed in positions required by the Standard
  • Sound level measurements may be taken outside the incubator for external alarm test
  • Results can be merged with vPad Safety Tests results and imported into commonly-used CMMS / Equipment Management Systems
  • Optional precision controlled heater permits testing of body temperature sensor accuracy

Standard Apps

vPad DOCS App Datrend Docs
Save, Store and Find Technical Documents.The vPad Datrend Docs is a technical document library, ideal for storing electronic manuals, procedures and specifications. Datrend Docs comes installed with all Operator Manuals for the vPad family.Compatible with:

  • vPad-IN
vPad-INR
Enables real-time testing of airflow, humidity, sound & temperatures.vPad-INR enables users to take measurements and capture data using all of vPad-IN sensors without waiting hours (IEC standards) for steady state temperature readings – ideal for when called into the ICU to diagnose a fault on an incubator.Compatible with:

  • vPad-IN
vPad-TMP
This App enables biomedical technicians to test medical thermometers, temperature probes, and temperature displays.Using the vPad-IN Oven, this App controls the oven, creates test records, and works independently from the vPad-IN system.Compatible with:

  • vPad-IN
vPad-TI
Automation App for testing Transporting IncubatorsCompatible with:

  • vPad-IN
vPad-RW
Automation App for testing Radiant WarmersCompatible with:

  • vPad-IN
Temperature:
Measurement Range: 0 deg C to 50 deg C
Accuracy: T1+/- 0.05 deg C
Resolution: 0.01 deg C
Convection: 5 “puck” sensors (T1-T5) – radiant warmer
Air Convection: 5 sensors (alternate “pos”’ T1 – T5) – incubator
1 Mattress Temperature Sensor
Skin Sensor2: Sensor “oven” with thermal pad contact
28° C to 40° C in 2° C increments
Resolution: 0.01° C
Accuracy: +/- 0.05° C
Airflow:
Measurement Range: 0.1 to 1 m/s
Accuracy: +/- 0.1 m/s at 50% RH +/- 15%
Sound Level:
Measurement Range: 30 dbA to 90 dbA, Type II
Accuracy: +/- 5 dbA
Resolution: 0.1 dbA
Relative Humidity:
Measurement Range: 0 – 100% Non-condensing
Accuracy: +/- 3 % to 100% RH
Resolution: 0.1% RH
Data Retention:
Capacity: up to 48 hours per test
Sample rate: 1/min or 1/sec, automatic, parameter dependent
Test Specifications:
Incubator Testing Specifications
  • Measures temperature at 6 points as per IEC 60601-2-19 (sub-clause 50.102)
  • Measures humidity at the center of the enclosure per IEC 60601-2-19 (sub-clause 103.1)
  • Measures airflow at 4 points in the enclosure per IEC 60601-2-19 (sub-clause 104.1)
  • Measures sound level inside the incubator per IEC 60601-2-19 (sub-clause 102.1)
  • Measures sound level inside the incubator during alarm per IEC 60601-2-19 (sub-clause 102.2)
  • Measures sound level outside the incubator per IEC 60601-2-19 (sub-clause 102.3)
  • Measures incubator temperature at center point (A) of incubator as per IEC 60601-2-19 (subclause 50.102)
  • Measures average incubator temperature (AIT) per IEC 60601-2-19 (sub-clause 2.9.106)
  • Measures incubator temperature (IT) per IEC 60601-2-19 (sub-clause 2.9.105)
  • Measures average temperature (AT) per IEC 60601-2-19 (subclause 2.9.103)
  • Measures average temperature at 4 points per IEC 60601-2-19 (sub-clause 50.102)
  • Measures skin temperature sensor accuracy per IEC 60601-2-19 (sub-clause 50.104)2
  • Confirms skin temperature sensor control of incubator per IEC 60601-2-19 (sub-clause 50.105)
  • Confirms independent incubator temperature sensor accuracy per IEC 60601-2-19 (sub-clause 50.106)
  • Measures difference between average temperature and control temperature per IEC 60601-2-19 (sub-clause 50.107)
  • Measures actual warm-up time compared to operating manual per IEC 60601-2-19 (sub-clause 50.108)
  • Measures overshoot and recovery to steady state per IEC 60601-2-19 (sub-clause 50.109)
  • Confirms humidity reading per IEC 60601-2-19 (sub-clause 50.110)

 

Radiant Warmer Testing Specifications
  • Measures skin temperature sensor accuracy per IEC 60601-2-21 (sub-clause 50.101)2
  • Measures average warmer temperature at 4 points per IEC 60601-2-21 (sub-clause 50.102)
  • Measures average warmer temperature at center point (A) of incubator as per IEC 60601-2-21(sub-clause 50.102)
  • Confirms skin temperature sensor control of warmer per IEC 60601-2-21 (sub-clause 50.103)
  • Confirms airflow is less than 0.1 m/s per IEC 60601-2-21 (subclause 50.102)
  • Measures sound level on the mattress per IEC 60601-2-21 (sub-clause 102.1)
  • Measures sound level on the mattress during alarm per IEC 60601-2-21 (sub-clause 102.1)
  • Measures sound level in front of the warmer per IEC 60601-2-21 (sub-clause 102.1)

All specifications subject to change without notice.

Please contact de Smit Medical Systems Ltd for further information about this product including;

  • Full product information / specification
  • Arranging a demonstration / evaluation
  • Pricing details
  • How to place an order
Call +44 (0)845 345 4226, or email us via our contact form.

Support Information

Helpdesk:
Head Office:
  • Address
    • de Smit Medical Systems Ltd, Unit 15 Falcons Gate, Dean Road, Bristol, BS37 5NH, UK
  • Operating Hours
    • Monday to Friday - 9am-5pm (Excluding bank holidays)
Service Centre:
  • See Head Office above
  • Please contact the Service Centre regarding any equipment servicing/repair related queries or telephone troubleshooting
Trained Personnel:
  • 7 x Sales technicians
  • 3 x Medical engineers at Head Office
    • All trained to manufacturer standard
Turnaround:
  • Warranty Repairs
    • 72 working hours response/resolver call-out for warranty repairs and service plans
  • Non-warranty Repairs
    • 72 hours subject to authorisation by Customer for paid repair
Spare Parts:
  • Spare parts for Mcube/Datrend products are available direct from de Smit Medical Systems Ltd
  • Next working day delivery if order received by 2pm on any given working day (subject to availability)

Two-Year Warranty on Datrend vPad Products

All Datrend vPad Products are designed and manufactured under an ISO 13485 registered Quality System and carry a one (1) year warranty from the date of original purchase.

This warranty will be automatically extended to two (2) years from the date of original purchase, provided that calibration is performed on an annual basis by de Smit Medical Systems.