Oxitest Plus 7

SpO2 Simulator – with Nellcor® Probe

Exclusive Nellcor-Puritan Bennett sensor port
The versatile, handheld Oxitest Plus7 provides the most advanced technology available in the market today for verifying the proper operation and performance of pulse oximeter monitors and modules throughout the clinical environment.

The Oxitest Plus 7 has an exclusive Nellcor-Puritan Bennett sensor port, which was developed in cooperation with Nellcor®. This Industry-exclusive feature allows correct selection of the instrument’s simulation output to the specific R-Cal value of the attached Nellcor® probe.

Compatibility beyond compare
To further enhance your pulse oximeter testing requirements, Oxitest Plus 7 is compatible with the ES601 Plus and other automated test systems such as the Fluke® Biomedical medTester 5000C.

Designed for compatibility with a wide variety of pulse oximeters, the Oxitest Plus 7 is the efficient, cost effective solution to your pulse oximeter testing requirements.

High Accuracy

Over 100 specific R-Curves

Multiple Pre-Set Tests

Add Sensitest

Analyse up to 5 unique probes

Easy To Use

Easily test SPO2 meters with exclusive built-in “finger” sensor

  • Easy adjustment of five preset oxygen saturation levels, six preset heart rates and four preset amplitudes.
  • Signal Artifact simulations
  • Preset patient parameters
  • Alarm Testing
  • Report printing
  • 2×20 character LCD display gives you all the information you need
  • Identifies the oximeter make and model
  • Status indicator of the test in progress with seven status messages
  • Rate
  • Saturation
  • Pulse Amplitude
Saturation (SpO2):
Range of adjustment is according to pulse oximeter manufacturer and/or model.
Masimo, Mindray, Nellcor, and Nellcor OEM’s (e.g., Critikon; Protocol; Philips; etc.):
Variable from 35% to 100% in steps of 1%
SpO2 presets at 35, 70, 80, 90, 93, and 97%
Selected EnviteC, Hewlett-Packard, Nihon Kohden, and Philips models:
Variable from 55% to 100% in steps of 1%
SpO2 presets at 55, 70, 80, 90, 93, and 97%
All other manufacturers and models:
Variable from 70% to 100% in steps of 1%
SpO2 presets at 70, 80, 90, 93, and 97%
SpO2 Accuracy:
All supported oximeters except Datex and Invivo oximeters:
55 to 100% SpO2: ±1% at specified SpO2 presets
35% SpO2: ±2%
Nellcor accuracy specified with either DS100A sensor, or with SpO2
Calibration selected via Nellcor Port
Datex and Invivo oximeters:
±1% at 97, 93% SpO2
±2% at 90, 80,
70% SpO2
Heart Rate:
Variable from 20 to 250 BPM in 1 BPM steps
Presets: 30, 60, 90,
120, 180, 240 BPM, Accuracy: ± 1 BPM
Pulse Amplitude:
Variable from zero (no blood flow) to 100% (normal adult pulse) in 1% steps.
Presets: 100%, 30%, 10%, and 5%, Accuracy: ± 1%
Signal Artifact:
Four preset simulations:
Movement
Tapping (Spike artifact)
Shivering
(Tremor artifact)
Shake Table (2.5Hz Sinewave)
Auto Presets:Nine preset patient simulations.
Default Auto Presets:
Normal Adult Tachycardia Low Perfusion
Hypoxia Bardycardia No perfusion
Movement Artifact Neonatal Tremor (Shivering Artifact)
Auto Preset parameters may be user-programmed via the RS-232 port.
Sensitest – Performance Specifications
Fault Conditions Displayed:
OPEN RED LED, INFRARED LED or PHOTODIODE LED or
PHOTODIODE SHORT to ground or shield LED SHORT to
PHOTODIODE Nellcor and Ohmeda connectors Standard.
Options: Up to 3 sensor connectors can be added to the enclosure
(total 5) from:
Datex, Hewlett Packard, Novamatrix, Spacelabs, Invivo, Datascope,
Criticare, Critikon, Sensormetics, PhysioControl, Simed and Nihon Koden.
Alarm Tests:
Five automated test sequences for determining oximeter alarm response time to Low SpO2 ; Low Heart Rate; High Heart Rate; Low Perfusion; and
Motion Artifact conditions.
Also determines time required for oximeter alarm to self-clear after test conditions are returned to normal.
Alarm Test parameters may be user-programmed via the RS-232 port.
Memory:
Auto Preset and Alarm Test results saved in battery-backed, non-volatile Memory, along with up to 20 PASS/FAIL records from manually-conducted tests. Test results in Memory may be printed as a Test Report via optional external Printer, or may be downloaded via RS-232 port to personal computer or automated tester.
User Interface:
Display: LCD – 20 character x 2 line
Keypad – 15 keys: PULSE OX, SPO (%), HEART RATE (preset), HR▲, HR▼, PULSE AMP (preset), PUL▲, PUL▼, PASS, FAIL, AUTO (preset), ALARM TEST, START/STOP (alarm test), ERASE (memory), PRINT
Nellcor Port:
Allows temporary connection of Nellcor pulse oximeter sensors for selection of sensor-specific SpO calibration curve. Intended for high-accuracy testing of all Nellcor disposable and reusable sensors.
Serial/Printer:
Mechanical: 5 Pin MiniDIN / Electrical:
Bi-directional
RS-232.
9600 baud, 8-N-1
Power Supply:
Battery: 7.2V NiCad rechargeable. Capacity: 1.4 Ah,
Battery life:
40 hours continuous
Standard Accessories:
Power Adapter (Region-specific)
Dimensions:
5.12 in x 7.75 in x 1.5 in (13 cm x 19.68 cm x 3.8 cm)
Weight:
1.93 lbs (875g)
Environment:
15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II

All specifications subject to change without notice.

Please contact de Smit Medical Systems Ltd for further information about this product including;

  • Full product information / specification
  • Arranging a demonstration / evaluation
  • Pricing details
  • How to place an order
Call +44 (0)845 345 4226, or email us via our contact form.

Support Information

Helpdesk:
Head Office:
  • Address
    • de Smit Medical Systems Ltd, Unit 15 Falcons Gate, Dean Road, Bristol, BS37 5NH, UK
  • Operating Hours
    • Monday to Friday - 9am-5pm (Excluding bank holidays)
Service Centre:
  • See Head Office above
  • Please contact the Service Centre regarding any equipment servicing/repair related queries or telephone troubleshooting
Trained Personnel:
  • 7 x Sales technicians
  • 3 x Medical engineers at Head Office
    • All trained to manufacturer standard
Turnaround:
  • Warranty Repairs
    • 72 working hours response/resolver call-out for warranty repairs and service plans
  • Non-warranty Repairs
    • 72 hours subject to authorisation by Customer for paid repair
Spare Parts:
  • Spare parts for Mcube/Datrend products are available direct from de Smit Medical Systems Ltd
  • Next working day delivery if order received by 2pm on any given working day (subject to availability)

Five-Year Warranty on Datrend legacy Products

All Datrend legacy products are designed and manufactured under an ISO13485 registered Quality System and a carry one (1) year from the date of original purchase. This warranty will be automatically extended on a yearly basis, to a maximum of five (5) years from the date of original purchase, provided that calibration is performed on an annual basis by de Smit Medical Systems.