vPad-BP, part of the vPad-A1 family, is a NIBP simulator designed to test the accuracy and repeatability of the NIBP function on Patient Monitors. User programmable settings allow for easy on site configuration to accommodate new models of NIBP devices.
- Compatible with all major industry manufacturers
- Field editable Envelopes allows for easy set up of new patient monitors
- Unlimited preset NIBP simulations
- Easy to navigate
- Auto Settings
- Auto Sequences
- Test reports
- Control via Bluetooth or USB
vPad-BP™ Performance Specifications
Manufacturer Envelopes | |
---|---|
|
Pressure Units | |
---|---|
mmHg, mbar, kPa, inH₂O, cmH₂O |
BP Presets | |
---|---|
Systolic/Diastolic (mmHg) |
Adult | Neonatal |
255/195 | 150/100 |
200/150 | 120/80 |
150/100 | 100/65 |
120/80 | 80/50 |
100/65 | 60/30 |
80/50 | 35/15 |
60/30 |
BP Simulation: | |
---|---|
Simulation type: | oscillometric |
Rate Range: | 20 – 240 BPM |
Accuracy: |
|
Amplitude: |
|
Amplitude Range: | 0 – 150% |
Amplitude Accuracy: | better than 0.5% |
AutoSettings: | unlimited, user definable |
Envelope Shift: | |
---|---|
± 50 mmHg max |
Minimum Diastolic: | 15 mmHg |
Maximum Systolic: | 275 mmHg |
Manometer: | |
---|---|
Pressure Range: | 0.0 to 400.0 mmHg |
Accuracy: | ± 0.5 mmHg |
Resolution: | 0.1 mmHg |
Regulated Pressure Source: | |
---|---|
Pressure Range: | 10.0 to 400.0 mmHg |
Accuracy: | ± 0.5 mmHg |
Resolution: | 0.1 mmHg |
Leak Test: | |
---|---|
Automatic/manual Inflation Automatic Timer |
Leak Test Time: | 30 – 600 seconds (user defined) |
Target Pressure: | 20 to 400 mmHg |
Range: | 0 to 200 mmHg/min |
User Definable Presets: | 12 |
User Definable AutoSettings: | unlimited |
OverPressure Test: | |
---|---|
Automatic/manual Inflation |
Range: | 20 – 400 mmHg |
Release Time: | 1-999 sec |
User Definable Presets: | 12 |
User Definable AutoSettings: | unlimited |
Standard Features / Accessories: | |
---|---|
|
Communication / User Interface | |
---|---|
via vPad-A1 Base Unit: |
Android 5″ tablet:
Dual XBUS for Datrend test automation |
Power Supply | |
---|---|
|
Dimensions | |
---|---|
98mm x 275mm x 97mm (3.87″ x 10.82″ x 3.80″) BP Unit (incl. A1 Base) |
Weight | |
---|---|
1080g (2.38lb) BP Unit (incl. A1 Base) 200g (0.44lb) wireless tablet interface |
Environment | |
---|---|
15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II |
All specifications subject to change without notice.
All Datrend vPad Products are designed and manufactured under an ISO13485 registered Quality System and carry a one (1) year warranty from the date of original purchase.
This warranty will automatically be extended to two (2) years from the date of original purchase, provided that calibration is performed on an annual basis by de Smit Medical Systems Ltd, exclusive UK Datrend Authorised Service Centre.