vPad-BP

NIBP Simulator

vPad-BP, part of the vPad-A1 family, is a NIBP simulator designed to test the accuracy and repeatability of the NIBP function on Patient Monitors. User programmable settings allow for easy on site configuration to accommodate new models of NIBP devices.

  • Compatible with all major industry manufacturers
  • Field editable Envelopes allows for easy set up of new patient monitors
  • Unlimited preset NIBP simulations
  • Easy to navigate
  • Auto Settings
  • Auto Sequences
  • Test reports
  • Control via Bluetooth or USB

vPad-BP™ Performance Specifications

Manufacturer Envelopes
  • GE Dinamap and Dinamap Pro
  • Critikon Dinamap Plus
  • Datascope Passport
  • Welch-Allyn Spot LXi and 52000
  • Fukuda Dynascope
  • Colin Press-Mate
  • …and more custom simulations
Pressure Units
mmHg, mbar, kPa, inH₂O, cmH₂O
BP Presets
Systolic/Diastolic (mmHg)
Adult Neonatal
255/195 150/100
200/150 120/80
150/100 100/65
120/80 80/50
100/65 60/30
80/50 35/15
60/30
BP Simulation:
Simulation type: oscillometric
Rate Range: 20 – 240 BPM
Accuracy:
  • ±0.25 BPM in sync mode
  • ±1 BPM otherwise
Amplitude:
  • 0 – 2 mL
  • 1.25 mmHg into 500ml cuff
Amplitude Range: 0 – 150%
Amplitude Accuracy: better than 0.5%
AutoSettings: unlimited, user definable
Envelope Shift:
± 50 mmHg max
Minimum Diastolic: 15 mmHg
Maximum Systolic: 275 mmHg
Manometer:
Pressure Range: 0.0 to 400.0 mmHg
Accuracy: ± 0.5 mmHg
Resolution: 0.1 mmHg
Regulated Pressure Source:
Pressure Range: 10.0 to 400.0 mmHg
Accuracy: ± 0.5 mmHg
Resolution: 0.1 mmHg
Leak Test:
Automatic/manual Inflation
Automatic Timer
Leak Test Time: 30 – 600 seconds (user defined)
Target Pressure: 20 to 400 mmHg
Range: 0 to 200 mmHg/min
User Definable Presets: 12
User Definable AutoSettings: unlimited
OverPressure Test:
Automatic/manual Inflation
Range: 20 – 400 mmHg
Release Time: 1-999 sec
User Definable Presets: 12
User Definable AutoSettings: unlimited
Standard Features / Accessories:
  • Autosequences
  • Unlimited User Defined Settings
  • vPad-A1 Power Base / Display
  • Universal Hose Adapter Kit
Communication / User Interface
via vPad-A1 Base Unit:

Android 5″ tablet:

  • Touchscreen User Interface
  • Wired (USB) or Bluetooth mode
  • WiFi
  • 16 GB memory

Dual XBUS for Datrend test automation

Power Supply
  • External AC adapter
  • Internal rechargeable Li-Ion batteries (for 10 hrs of simulation with full charge)
Dimensions
98mm x 275mm x 97mm (3.87″ x 10.82″ x 3.80″)
BP Unit (incl. A1 Base)
Weight
1080g (2.38lb) BP Unit (incl. A1 Base)
200g (0.44lb) wireless tablet interface
Environment
15°C to 40°C, 10% to 90% RH, Indoor Use Only,
Category II

All specifications subject to change without notice.

Two-Year Warranty on Datrend vPad Products

All Datrend vPad Products are designed and manufactured under an ISO 13485 registered Quality System and carry a one (1) year warranty from the date of original purchase.

This warranty will be automatically extended to two (2) years from the date of original purchase, provided that calibration is performed on an annual basis by de Smit Medical Systems.

The warranty on Android devices is one year and cannot be extended.

vPad-A1 - Introduction