vPad-O2 is the next generation SpO2 tester. Built on the vPad platform, vPad-O2 is user programmable so new sensors can easily be added in the field without needing to send the unit back to a service centre for upgrading. Being part of the vPad-A1 family, vPad-O2 can be upgraded to include a Multi-Parameter Patient Simulator and / or a NIBP Simulator.
- Compatible with all major industry manufacturers
- Field editable R Curves allows for easy set up of new oximeters
- Unlimited preset patient simulations
- Five automated test sequences for determining oximeter alarm response time
- Four preset signal artifact simulations
- Easy to navigate
- Control via Bluetooth or USB
- Auto Settings
- Auto Sequences
- Test reports
- Field upgradeable
vPad-O2™ Performance Specifications
Saturation (SpO₂) | |
---|---|
Range: | 30% to 100% |
Increments: | 1% |
Presets: | 6, user definable |
Range of adjustment and presets may vary according to pulse oximeter specifications |
SpO₂ Accuracy | |
---|---|
Saturation within DUT specified range: | ±1 count + specified accuracy of the DUT |
Heart Rate | |
---|---|
Range: | 20 to 300 BPM |
Increments: | 1 BPM |
Presets: | 6, user definable |
Accuracy: | ± 0.25 BPM (sync mode); otherwise, ±1 BPM |
Pulse Amplitude | |
---|---|
Range: | 0 to 100% |
Increments: | 1% steps. |
Presets: | 6, user definable |
Accuracy: | ± 1% |
Signal Artifact | |
---|---|
Four preset simulations: |
|
Auto Presets: | |
---|---|
Unlimited preset patient simulations |
Default Auto Presets: |
|
Alarm Tests: | |
---|---|
Automated test sequences for determining oximeter alarm response time to: | |
|
|
Five defaults, plus unlimited programmable alarm sequences |
Communication / User Interface | |
---|---|
via vPad-A1 Base Unit: |
Android 5″ tablet:
Dual XBUS for Datrend test automation |
Power Supply | |
---|---|
via vPad-A1 Base Unit: |
|
Dimensions | |
---|---|
98mm x 208mm x 30mm (3.85” x 8.2” x 1.18”) A1 Base 90mm x 160mm x 24mm (3.54” x 6.3” x 0.95”) SpO₂ Probe |
Weight | |
---|---|
440g (0.96lb) A1 base 122g (0.27lb) SpO₂ Probe |
Environment | |
---|---|
15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II |